Table 2 The roughness ratio, detection sensitivity, Imax and K m obtained from experiments

From: Role of reaction kinetics and mass transport in glucose sensing with nanopillar array electrodes

Specimen Roughness Ratio Sensitivity of bare electrodes to K4Fe(CN)6 (μA·mM-1·cm-2) Sensitivity of functionalize electrodes to glucose (μA·mM-1·cm-2) Imax glucose (μA) K m glucose (mM)
Flat 1.0 19.30 0.27 1.34 24.8
Nano A 20.0 41.40 0.91 5.06 29.3
Nano B 38.8 41.05 1.80 10.1 32.6
Nano C 63.4 41.70 3.13 23.0 52.0