Figure 6 | Journal of Biological Engineering

Figure 6

From: Development of a modified Respiration Activity Monitoring System for accurate and highly resolved measurement of respiration activity in shake flask fermentations

Figure 6

Comparison of the method of Anderlei et al.[23, 24]using equation (A.2) ( blue triangle ) and the new evaluation method using equation (B.1) in the rinsing phase ( open square ) and equation (B.4) in the stop phase ( black square ). A: Enlarged section of the OTR curve of Figure5A; B: E. coli BL21 pRSET eYFP-IL6 in TB medium; temperature T = 30°C; filling volume VL = 10 mL; shaking diameter d0 = 50 mm; shaking frequency n = 300 rpm; inoculated with 0.2 ml cryo culture; rinsing phase trinse = 55 min; stop phase tstop = 5 min; high flow phase thigh = 0.9 min; C: Gluconobacter oxydans 621 H wildtype in complex medium with 40 g/L mannitol and 0.5 g/L glycerol; temperature T = 30°C; filling volume VL = 10 mL; shaking diameter d0 = 50 mm; shaking frequency n = 350 rpm; initial optical density OD600 = 0.1; rinsing phase trinse = 25 min; stop phase tstop = 5 min; a sensor lag time of τ = 0.013 h is used for all experiments.