Table 1 The accuracy of FP-based quantification method with HT-29 EVs

From: A simple approach for rapid and cost-effective quantification of extracellular vesicles using a fluorescence polarization technique

Sample Added EV counts (× 107) Measured EV counts (× 107) SDa CV (%)b Recovery (%)c
A 248 247 18.7 7.6 99.7
B 360 343 12.4 3.6 95.3
C 550 558 54.2 9.7 101.5
D 760 765 3.6 0.5 100.7
  1. aStandard deviation of three measurements
  2. bCoefficient of variation (%) = SD/mean × 100
  3. cRecovery (%) = Measured value/added value × 100