Future developments in biosensors for field-ready Zika virus diagnostics

  • Ariana M. Nicolini1,
  • Katherine E. McCracken2 and
  • Jeong-Yeol Yoon1, 2Email author
Contributed equally
Journal of Biological Engineering201711:7

https://doi.org/10.1186/s13036-016-0046-z

Received: 3 November 2016

Accepted: 27 December 2016

Published: 23 January 2017

Abstract

Since early reports of the recent Zika virus outbreak in May 2015, much has been learned and discussed regarding Zika virus infection and transmission. However, many opportunities still remain for translating these findings into field-ready sensors and diagnostics. In this brief review, we discuss current diagnostic methods, consider the prospects of translating other flavivirus biosensors directly to Zika virus sensing, and look toward the future developments needed for high-sensitivity and high-specificity biosensors to come.

Keywords

Zika Flaviviruses Biosensors RT-PCR LAMP Immunoassays

Background

Amidst the recent Zika epidemic, rising public health concerns have led to extensive research aimed at uncovering the underlying mechanisms of Zika virus (ZIKV) infection and transmission pathways [13]. According to the Pan American Health Organization (PAHO), autochthonous ZIKV cases in the Americas increased from virtually none in early 2015 to over 170,000 confirmed and 515,000 suspected cases by December 2016 [4]. This escalation has led to newly abundant clinical, epidemiological, and virological research and funding opportunities that were previously limited by the rarity of infection and limited concerns for ZIKV as an infectious agent (Fig. 1). Interestingly, research aimed at developing novel ZIKV sensors is quite limited, as seen in Fig. 1. A whole arena remains open for research, funding, and commercial opportunities.
Fig. 1

Number of peer-reviewed publications on ZIKV related to novel sensors development, topic reviews and commentaries, molecular biology and virology, and epidemiology or clinical evaluations of Zika cases (as of October 15, 2016). Cumulative publications are presented in 5 year increments until 2015 and 1 year increments between 2015 and 2016 (top). Publications in 2015–2016 are also presented separately by month (bottom)

Between its first isolation as a zoonotic pathogen in Uganda (1947) and the first major human outbreak of Zika on Yap Island in Micronesia (2007), ZIKV has been primarily observed in Africa and the Pacific [57]. Generally, the flu-like symptoms of infection are mild and include low to moderate fever, headache, joint pain, rash, and fatigue [610]. However, the recent breadth of epidemiological data stemming from the many thousands of cases across South America, the Caribbean, and Central and North America, has uncovered new insights into rare and severe effects on specific subsets of the population. These include a low risk of Guillain-Barré syndrome in adults, and critical risks for pregnant women, including stillbirth, restricted intrauterine fetal growth, and microcephaly [7, 1014].

As a member of the Flavivirus genus, ZIKV shares many common genetic sequences and protein structures with other high-interest flaviviruses, including Dengue virus (DENV), West Nile virus (WNV), yellow fever virus (YFV), and Spondweni virus, its most similar relative [15, 16]. On a molecular level, ZIKV features a 10.7 kb single-stranded and positive-sense RNA genome. The polyprotein that this genome encodes cleaves to form several structural proteins, including the envelope (E) and membrane (M) proteins, and nonstructural (NS1 and NS5) proteins [17]. These proteins are the common focus in immunosensing and molecular research for other flaviviruses [12, 13, 1618]. Thus, despite the historically limited attention given to ZIKV in the research community, previous work with other flaviviruses may help to inform a rapid turnaround in future ZIKV sensing technologies [7, 8, 15].

In the shadow of the recent epidemic, our understanding of ZIKV pathogenicity has expanded at both the population level and the molecular level. Although several benchtop methods for ZIKV detection have been employed for emergency use, there is still a need for the development and funding of alternative field-ready diagnostic tools. Prompt identification of ZIKV infection at the site of exposure from patient-direct samples is critical in minimizing the global spread of the virus. During the ongoing development and rapid expansion of the ZIKV sensor market, target specificity and sensitivity amidst complex sample matrices are key. In this brief review, we highlight current techniques, emergent diagnostic methods, and considerations for developing future field-ready biosensors.

Gold standards of ZIKV detection

The recent increase in the number of ZIKV cases, particularly in the U.S., has led the U.S. Food and Drug Administration (FDA) to issue an Emergency Use Authorization (EUA) for several previously non-cleared or unapproved diagnostic assays. The FDA and U.S. Centers for Disease Control and Prevention (CDC) have recommended that ZIKV detection in human patients be performed by reverse transcription quantitative real-time polymerase chain reaction (RT-qPCR), or by serological tests using an IgM antibody capture enzyme-linked immunosorbent assay (MAC-ELISA) or a plaque-reduction neutralization test (PRNT) (Fig. 2) [19].
Fig. 2

Flow diagram of gold standard ZIKV detection by patient type and time from onset of symptoms. RT-qPCR detection is typically used in the first 2 weeks of illness and IgM ELISA after the first 2 weeks or when RT-qPCR is negative. PRNT assay should be used as a final test if an ELISA assay returns positive or inconclusive

For asymptomatic pregnant women who have traveled to high-risk areas for ZIKV and for symptomatic individuals within the first 2 weeks of symptom onset, the preferred detection method authorized by the FDA EUA is the Trioplex RT-qPCR assay, which is specific to DENV, Chikungunya virus (CHIKV), and ZIKV. In RT-qPCR, a patient sample is added to a buffered reagent solution containing target primers, reverse transcriptase (to generate cDNA from viral RNA), DNA polymerase (to amplify this cDNA), deoxynucleotides (dNTPs), and an intercalating fluorescent dye or fluorescent reporter (Fig. 3a). The amplified target is then quantified by absolute or relative fluorescence following a given number of thermocycles, typically lasting 90–120 min. This assay can be performed in the presence of several sample matrices including serum, whole blood, cerebrospinal fluid, urine, and amniotic fluid [20]. Although RT-PCR is inherently very sensitive, the possibility of false-negatives is high. Therefore, testing of symptomatic patients with negative RT-PCR results should be confirmed with alternative forms of identification.
Fig. 3

Assay schematics for ZIKV diagnostics by reverse transcription quantitative real-time polymerase chain reaction (RT-qPCR), IgM antibody capture enzyme-linked immunosorbent assay (MAC-ELISA), and a plaque-reduction neutralization test (PRNT). a In one-step RT-qPCR, a patient sample is thermally cycled in a buffered reagent solution containing ZIKV primers, and the amplified target is identified by fluorescence, typically after 40 cycles. b In MAC-ELISA, human IgM developed in response to ZIKV infection are captured and quantified though antibody interactions and enzymatic conversion of a chromogenic substrate. c In PRNT, patient serum dilutions are mixed with live virus samples and are applied to confluent host cells. Antibodies in infected patients neutralize the virus, leading to a reduction in observable plaques

For later stages of ZIKV infection, antibody-based methods can be used. Typically, neutralizing antibodies to ZIKV develop in the human body within the first week of symptoms and continue to remain at detectable levels for up to 12 weeks. During this timeframe, serological assays can be performed to detect the patient’s anti-ZIKV IgM antibodies. However, due to the epidemiological and molecular similarities of ZIKV to other flaviviruses, IgM ELISA assays should be conducted for the antibodies formed against ZIKV, DENV, and CHIKV. In MAC-ELISA, a patient’s sample is added to a well plate pre-coated with antibodies to capture human IgM (Fig. 3b). A virus-specific antigen is then added and washed away, binding specifically to the IgM of infected patients. Finally, an antibody specific to this same viral target that is tagged with an enzyme (e.g. horseradish peroxidase) is added and a chromogenic substrate is used for quantification. Samples from infected patients will thus elicit an optically detectible signal (e.g. absorbance, fluorescence) that may be correlated to IgM concentration. However, the risk of false-positives is high for IgM and IgG assays. If ELISA testing is inconclusive or positive, PRNT should be performed to confirm the presence of ZIKV, specifically [18].

PRNT typically serves as a secondary test to IgM ELISA and measures the ability of a patient’s antibodies to neutralize a specific virus. In PRNT, serial dilutions of a patient’s serum are added to samples of a viral suspension, and each mixture is applied to a confluent host cell culture (e.g. Vero cells) (Fig. 3c). Following incubation, plaque forming units (PFU) are counted. If neutralizing antibodies specific to this virus are present in the patient’s serum, the associated PFU value will be reduced, and the antibody titer can be determined from the serial dilutions. This method provides better sensitivity and specificity over IgM ELISA, but requires extended time (days), labor, materials, and therefore cost.

Developmental diagnostic methods

Outside of tests that the CDC offers, there are several private companies selling RT-PCR, ELISA, and lateral flow assay kits (Table 1). Many research groups have also focused on alternative sensing modalities that reduce extensive sample preparation, the use of expensive laboratory equipment, and the risks of false-positive and false-negative results characteristic of ELISA and PCR assays. Some of these recent research findings are based on techniques previously used for the detection of other flaviviruses, whereas others are novel sensors unique to ZIKV.
Table 1

Laboratory-based ZIKA assay kits

Assay technique

Manufacturer

Product name

ELISA

Alpha Diagnostic International

Recombivirus Human Anti-Zika Virus (ZIKV) Envelope protein IgG/IgM ELISA kits

CDC

aTrioplex Real-time RT-PCR Assay

CTK Biotech

RecombiLISA Zika IgM ELISA kit

DIA.PRO Diagnostic Bioprobes

ZIKV IgG/M ELISA kits

Euroimmun

Anti-Zika Virus IgG/IgM ELISA kits

InBios International

aZIKV Detect IgM Capture ELISA kit

MyBioSource

Zika Virus IgM (ZV-IgM) ELISA kit

NovaTec Immundiagnostica

NovaLisa Zika Virus IgM μ-capture ELISA kit

Lateral flow assay

Chembio Diagnostic Systems

DPP Zika IgM/IgG assay

Immunofluorescence

Euroimmun

Anti-Zika Virus IIFT (IgG or IgM)

Multiplex RT-qPCR

Bioneer

AccuPower® ZIKV (DEN, CHIKV) Multiplex Real-time RT-PCR Kit

Luminex

azMAP® MultiFLEXTM Zika RNA Assay

SolGent

DiaPlexQTM ZCD Detection Kit

ThermoFisher Scientific

TaqMan Arbovirus Triplex Assay (ZIKV/DENV/CHIKV)

RT-qPCR

Altona Diagnostics

aRealStar Zika Virus RT-PCR Kit 1.0

BioinGentech

HumqPCR-realtimeTM Zika Detection

CDC

Trioplex Real-time RT-PCR Assay

Coyote Bioscience

One Step qPCR Detection kit for the Zika Virus

DaAn Gene

Detection Kit for Zika Virus RNA

Focus Diagnostics

aZika Virus Qualitative Real-Time RT-PCR

Liferiver

Zika Virus Real Time RT-PCR

MyBioSource

Zika, PCR Kit

Roche Molecular Systems

aLightMix® Zika rRT-PCR Test

Siemens Healthcare Diagnostics

aVERSANT® Zika RNA 1.0. Assay (kPCR)

ThermoFisher Scientific

TaqMan Zika Virus Singleplex Assay

US Biological Life Sciences

Genesig Teal-Time PCR Kit for Zika Virus, Easy

Viracor-IBT Laboratories

aZika Virus Real-time RT-PCR Test

WELLS BIO

careGENETM Zika Virus RT-PCR Kit

RT-PCR

ARUP

aZika Virus Detection by RT-PCR Test

Vela Diagnostics USA

a Sentosa® SA ZIKV RT-PCR Test

Transcription-mediated amplification

Hologic

aAptima® Zika Virus Assay

aIndicates assays approved under the FDA Emergency Use Authorization (as of October 21, 2016)

Molecular detection of ZIKV nucleic acid

Reverse transcription PCR (RT-PCR) has become the gold standard for molecular amplification and detection of viruses because of its high selectivity and relatively high sensitivity. Prompted by the 2007 ZIKV outbreak in Yap State, Micronesia, several RT-PCR methods have been developed to specifically identify a multitude of ZIKV strains independent of other flaviviruses. Published ZIKV-specific primer sets target highly conserved regions of the structural membrane (M) and/or envelope (E) [21] proteins, partial envelope (pE) [9] protein, or the non-structural (NS1 and NS5) proteins [18, 2225]. Detection of ZIKV using RT-PCR has also been shown to work in the presence of many sample matrices including plasma [26], serum [21, 27], saliva [28], urine [27], conjunctival fluid, and semen [29], thus reducing the need for sample purification or extraction.

Many commercial nucleic acid amplification tests (NAATs) have been developed for ZIKV detection within the past year (Table 1). Between February 26, 2016 and October 21, 2016, the FDA approved ten molecular diagnostic tests for clinical identification of ZIKV under the EUA [30]. Eight of these assays utilize traditional RT-PCR or RT-qPCR amplification (conventional and quantitative real-time thermal cycling) and detection (gel electrophoresis or intercalating dye fluorescence intensity) methods.

The following two non-traditional FDA EUA-approved NAATs claim improved sensitivity, specificity, usability, and speed. The xMAP® MultiFLEXTM assay (Luminex Corp.), uses a proprietary device to complete a series of steps, which include RT-PCR, followed by amplicon-particle hybridization, and final detection via an indicator molecule [31]. The other, Aptima Zika virus assay (Hologic, Inc.), also uses a proprietary device; however, this assay is fully automated and can perform transcription-mediated amplification (exact technique not specified), and qualitative viral detection in the presence of human serum, plasma, or urine, similar to the xMAP® MultiFLEX™ assay [32]. Despite claiming ease-of-use and rapid sample-to-answer times, both methods require approximately 3.5 h and expensive laboratory equipment, and thus laboratory space.

In the case of epidemic diseases, extremely rapid and low-cost screening of clinical samples in-field is necessary, making these EUA techniques inadequate. In light of this need, many research groups have focused on making PCR assays field-deployable and/or field-ready [3337]. Although some have succeeded in creating full sample-to-answer devices (Fig. 4a), PCR platforms are still limited by their need for multi-temperature sample heating for denaturation, annealing and extension. Fortunately, over the past 30 years isothermal amplification techniques with typical amplification times of less than 1 h have been thoroughly described for a variety of DNA and RNA targets. Popular forms of isothermal NAATs include nucleic acid sequence-based amplification (NASBA), loop-mediated isothermal amplification (LAMP), strand invasion based amplification (SIBA), strand displacement amplification (SDA), helicase-dependent amplification (HAD), recombinase polymerase amplification (RPA) and others [38].
Fig. 4

a Palm-sized device for point-of-care Ebola detection using RT-PCR and fluorescence detection (reproduced from ref. 33 with permission from American Chemical Society). b Lab-on-a-CD integrated LAMP for foodborne pathogen detection (reproduced from ref. 45 with permission from Elsevier). c Instrument-free RT-LAMP assay and self-contained cassette for point-of-care ZIKV assay (reproduced from ref. 40 with permission from American Chemical Society)

Since the first publication by Pardee et al. in May 2016, four groups have published research on isothermal-NAAT ZIKV detection using NASBA [39], RT-LAMP [40, 41] and RT-SIBA [42] (Table 2), several of which are still laboratory-based. All four groups also used different amplicon detection modalities, including toehold switch sensors, colorimetric detection, AC susceptometry, and gel electrophoresis. The RT-LAMP assay developed by Song et al. is particularly noteworthy due to its self-contained and field-ready design, which allows identification of ZIKV in under an hour on a portable cassette for less than $2 per assay (Fig. 4c) [40].
Table 2

ZIKV biosensors developed in 2016

Assay technique

Detection mode

Target

Sample matrix

Range of detection

Limit of detection

Sample volume

Assay time

Cost per assay

Ref.

Immunoassay

Impedimetry

NS1 (flaviviruses)

PBS

Serum

10-2000 ng/mL

10-1000 ng/mL

3 ng/mL

30 ng/mL

40 μL

30 min

unspecified

[61]

Capacitance

NS1 (flaviviruses)

PBS

Serum

5-1000 ng/mL

5-1000 ng/mL

0.2 ng/mL

0.5 ng/mL

Immunoassay

Chemiluminescence

E protein (ZIKV)

PBS

Urine, plasma

10-105 PFU/mL

10-104 PFU/mL

10 PFU/mL

10 PFU/mL

100 μL

~2 h

unspecified

[68]

LAMP

AC susceptometry

NS5 oligonucleotide (ZIKV)

Unspecified

Serum

1-103 aM

1-104 aM

1 aM

1 aM

40 μL

<30 min

unspecified

[41]

NASBA-CRISPR

Colorimetry

RNA (ZIKV)

Serum

3 fM - 30 pM

1 fM

300 μL

~3 h

$0.10-$1

[39]

RT-LAMP

Colorimetry

E protein RNA (ZIKV)

Saliva

50-5 × 104 PFU/mL

50-100 PFU/mL

65 μL

40 min

~$2

[40]

RT-SIBA

Fluorescence

RNA (ZIKV)

Lysis buffer

5 × 103-5 × 106 copies/mL

5000 copies/mL

2 μL

<30 min

unspecified

[42]

Of the molecular diagnostic techniques, isothermal genomic amplification has arguably become the most promising method for in-field pathogen identification due to its enhanced specificity, decreased limit-of-detection, reduced assay time, ease of amplification, and number of end-product detection methods. Although not all for ZIKV, several groups have already developed low-cost devices using inexpensive insulating materials (e.g. thermoses) [40, 43, 44] and simple heat-producing elements [4548], including non-electrical exothermal reactions [43, 49, 50] (Fig. 4b,c). Many real-time nucleic acid quantification methods have also been used, though again not all for ZIKV, and include measurement of fluorescence [51], Mg+ pyrophosphate [52], electrochemical [53], or colorimetric signal changes, detectable by the human eye [54, 55] or optical sensors [5658].

Antibody-based assays

Despite advances in molecular diagnostics, the cost of reagents and equipment and the likelihood of false-negative results present inherent challenges. For these reasons, serological assays remain important alternatives or supplements for detection, particularly when focusing on field-ready assays [18]. These techniques have been most commonly used to detect a patient’s antibody response within a diseased state, as previously described, but can also be extended to direct assays for ZIKV antigens in any sample matrix (i.e. immunoassay), including mosquito pool samples. Flavivirus immunoassays, including ELISA and antibody-based lateral flow assays, have primarily been developed through the antibodies to NS1, NS5, or E proteins [59]. These are also the principal routes of detection in commercial ZIKV MAC-ELISA kits authorized by the FDA (Table 1) [60].

Depending on the extent of conservation for a targeted epitope among all flaviviruses, some existing assays for DENV or YFV may be adapted directly to ZIKV, but will only have the resolution to broadly identify ZIKV as a flavivirus. Recent immunosensors developed and tested by Cecchetto et al., for example, use impedimetric and capacitive sensing of the NS1 protein from DENV, and have the potential for near immediate conversion to ZIKV detection due to potential cross-reactivity of the anti-NS1 IgG1 antibodies employed [61] (Table 2). Similarly, flavivirus biosensors have been developed using lab-on-a-chip and lab-on-a-CD technology for label-free optical and electrochemical sensing of DENV through serological IgM or NS1 protein binding [62, 63].

Because ZIKV and other flaviviruses are similarly transmitted to humans by mosquitoes of the Aedes genus, though, the origin of a biomarker detected by a nonspecific immunoassay may be unclear [64]. This potential for cross-reactivity is a primary concern for ZIKV immunosensing and begs further research into high-affinity antibodies with greater species specificity.

In response to these concerns, new research by Dai et al. has focused on discerning the operation of flavivirus antibody recognition for ZIKV through improved characterization of surface protein structures at the angstrom level [65]. Their work has discerned one mode of antibody binding specifically to the ZIKV E protein along a conserved fusion loop, which may be a focal point for future targeted sensors. An extensive survey of E protein structures across 50 ZIKV strains by Badawi et al. has also confirmed multiple conserved epitopes between these, and work by Zhao et al. has revealed several candidate mouse antibodies that demonstrate favorable specificity for ZIKV detection through binding localization at the DIII feature of the ZIKV E protein [59, 66]. However, other proteins may also be ideal candidates for sensing methods. For example, Meltzer et al. have recently highlighted the merits of developing IgM and IgG that are specific to the ZIKV NS1 protein, through which detection may also be more species-specific [67].

Following these efforts, early steps toward instrument-free and point-of-care (i.e. field-ready) ZIKV-specific immunosensors have been reported, although these have been few in number. For example, Acharya et al. developed a chemiluminescent immunoassay that specifically detects ZIKV by recognition of the E protein and quantification following magnetic particle separation and immunoblotting (Fig. 5) [68]. Exploration into immunosensing methods adaptable to field-ready diagnostics promises substantial improvements in future ZIKV detection and treatment, especially if cross-reactivity can be eliminated with new high-affinity, high-specificity antibodies.
Fig. 5

a Biocan diagnostic’s Tell Me Fast™ Zika/Dengue/Chikugunya virus IgG/IgM lateral flow assay (reproduced from www.zikatest.com with permission from Biocan Diagnostics, Inc.). b Chemiluminescent particle immunoassay for ZIKV detection by magnetic separation and ultraviolet fluorescence (reproduced from ref. 68 with permission from the authors)

Conclusions

Considerable research is still required to reach our goals of ZIKV sensing across an array of sample matrices with field-ready assay platforms. Fortunately, much has been learned over the past year about ZIKV on a molecular level, and thus many new opportunities have emerged for applying this knowledge toward treatment and diagnostics. Molecular identification of not only ZIKV, but also of other flaviviruses, hinges on the implementation of alternative techniques for amplicon production and detection. Strides have been made in the design of suitable primer sets specific to flaviviruses and ZIKV in particular; however, more development is needed for rapid, in-field detection. For antibody-based assay development, researchers may build their methods from existing cross-reactive assays, but adoption of forthcoming ZIKV-specific antibodies will be necessary for improved specificity. Above all, sensors that can be quickly and inexpensively assembled, screened for quality, and deployed will make the greatest impact in helping to understand and prevent the spread of ZIKV.

Abbreviations

CDC: 

U.S. Centers for Disease Control and Prevention

CHIKV: 

Chikungunya virus

DENV: 

Dengue virus

EUA: 

Emergency Use Authorization

FDA: 

U.S. Food and Drug Administration

HAD: 

Helicase-dependent amplification

LAMP: 

Loop-mediated isothermal amplification

MAC-ELISA: 

IgM antibody capture enzyme-linked immunosorbent assay

NAATs: 

Nucleic acid amplification tests

NASBA: 

Nucleic acid sequence based amplification

PRNT: 

Plaque-reduction neutralization testing

RPA: 

Recombinase polymerase amplification

RT-PCR: 

Reverse transcription polymerase chain reaction

RT-qPCR: 

Reverse transcription quantitative real-time polymerase chain reaction

SDA: 

Strand displacement amplification

SIBA: 

Reverse transcription-strand invasion based amplification

WNV: 

West Nile virus

YFV: 

Yellow fever virus

ZIKV: 

Zika virus

Declarations

Acknowledgements

Not applicable.

Funding

Funding for this research was provided by BIO5 Institute, The University of Arizona. AMN acknowledges the support of the SMART scholarship from U.S. Department of Defense. KEM acknowledges support from the U.S. National Science Foundation Graduate Research Fellowship under grant number DGE-1143953. Any opinion, findings, and conclusions or recommendations expressed in this material are those of the author(s) and do not necessarily reflect the views of the National Science Foundation.

Availability of data and materials

Data sharing is not applicable to this article as no primary datasets were generated or analyzed in this work.

Authors’ contributions

AMN and KEM jointly prepared the manuscript and figures with consultation and guidance from JYY. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Figure 4a is reproduced from ref. 33 with permission from the American Chemical Society. Figure 4b is reproduced from ref. 45 with permission from Elsevier. Figure 4c is reproduced from ref. 40 with permission from American Chemical Society. Figure 5a is reproduced from www.zikatest.com with permission from Biocan Diagnostics, Inc. Figure 5b is reproduced from ref. 68 with permission from the authors.

Ethics approval and consent to participate

The need for ethics approval and consent was waived.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
Biomedical Engineering Graduate Interdisciplinary Program and Department of Biomedical Engineering, The University of Arizona
(2)
Department of Agricultural and Biosystems Engineering, The University of Arizona

References

  1. Hamel R, Liégeois F, Wichit S, Pompon J, Diop F, Talignani L, et al. Zika virus: epidemiology, clinical features and host-virus interactions. Microbes Infect. 2016;18:441–9.View ArticleGoogle Scholar
  2. Chang C, Ortiz K, Ansari A, Gerschwin ME. The Zika outbreak of the 21st century. J Autoimmun. 2016;68:1–13.View ArticleGoogle Scholar
  3. Weaver SC, Costa F, Garcia-Blanco M, Ko AI, Ribeiro GS, Saade G, Shi P-Y, Vasilakis N. Zika virus: History, emergence, biology, and prospects for control. Antivir Res. 2016;130:69–80.View ArticleGoogle Scholar
  4. Pan American Health Organization (PAHO) / World Health Organization (WHO). Zika Epidemiological Update — 15 December 2016. PAHO/WHO. 15 Dec 2016Google Scholar
  5. Dick GW, Kitchen SF, Haddow AJ. Zika virus (I). Isolations and serological specificity. T Roy Soc Trop Med H. 1952;46:509–20.View ArticleGoogle Scholar
  6. Duffy MR, Chen T-H, Hancock WT, Powers AM, Kool JL, Lanciotti RS, et al. Zika virus outbreak on Yap Island, federated states of Micronesia. N Engl J Med. 2009;360:2536–43.View ArticleGoogle Scholar
  7. Posen HJ, Keystone JS, Gubbay JB, Morris SK. Epidemiology of Zika virus, 1947–2007. BMJ Glob Health. 2016;1:e000087.View ArticleGoogle Scholar
  8. Kindhauser MK, Allen T, Frank V, Santhana RS, Dye C. Zika: the origin and spread of a mosquito-borne virus. Bull World Health Organ. 2016;94:675–686C.View ArticleGoogle Scholar
  9. Lanciotti RS, Kosoy OL, Laven JJ, Velez JO, Lambert AJ, Johnson AJ, et al. Genetic and serologic properties of Zika virus associated with an epidemic, Yap State, Micronesia, 2007. Emerg Infect Dis. 2008;8:1232–9.View ArticleGoogle Scholar
  10. Shuaib W, Stanazai H, Abazid AG, Mattar AA. Re-emergence of Zika virus: a review on pathogenesis, clinical manifestations, diagnosis, treatment, and prevention. Am J Med. 2016;129:879.e8-11.View ArticleGoogle Scholar
  11. Rasmussen SA, Jamieson DJ, Honein MA, Petersen LR. Zika virus and birth defects — reviewing the evidence for causality. N Eng J Med. 2016;374:1981–7.View ArticleGoogle Scholar
  12. Brasil P, Pereira JP, Gabaglia CR, Damasceno L, Wakimoto M, Nogueira RMR, et al. Zika virus infection in pregnant women in Rio de Janeiro - preliminary report. N Eng J Med. 2016. doi:https://doi.org/10.1056/NEJMoa1602412
  13. Faria NR, da Silva Azevedo RDS, Kraemer MUG, Souza R, Cunha MS, Hill SC, et al. Zika virus in the Americas: early epidemiological and genetic findings. Science. 2016;352:345–9.View ArticleGoogle Scholar
  14. Aragao MDFV, van der Linden V, Brainer-Lima AM, Coeli RR, Rocha MA, da Silva PS, et al. Clinical features and neuroimaging (CT and MRI) findings in presumed Zika virus related congenital infection and microcephaly: retrospective case series study. Brit J Med. 2016;353:i1901.View ArticleGoogle Scholar
  15. Bollati M, Alvarez K, Assenberg R, Baronti C, Canard B, Cook S, et al. Structure and functionality in flavivirus NS-proteins: perspectives for drug design. Antivir Res. 2010;87:125–48.View ArticleGoogle Scholar
  16. Barba-Spaeth G, Dejnirattisai W, Rouvinski A, Vaney M-C, Medits I, Sharma A, et al. Structural basis of potent Zika-dengue virus antibody cross-neutralization. Nature. 2016;536:48–53.View ArticleGoogle Scholar
  17. Lindenbach BD, Rice CM. Molecular biology of flaviviruses. Adv Virus Res. 2003;59:23–61.View ArticleGoogle Scholar
  18. Musso D, Gubler DJ. Zika virus. Clin Microbiol Rev. 2016;29:487–524.View ArticleGoogle Scholar
  19. CDC (U.S. Centers for Disease Control and Prevention). Types of Zika Virus Tests. CDC. 7 Sept 2016. http://www.cdc.gov/zika/laboratories/types-of-tests.html. Accessed 2 Oct 2016.
  20. CDC. Trioplex Real-time RT-PCR Assay. CDC. 21 Sept 2016. http://www.cdc.gov/zika/pdfs/trioplex-real-time-rt-pcr-assay-instructions-for-use.pdf. Accessed 2 Oct 2016.
  21. Faye O, Faye O, Dupressoir A, Weidmann M, Ndiaye M, Sall AA. One-step RT-PCR for detection of Zika virus. J Clin Virol. 2008;43:96–101.View ArticleGoogle Scholar
  22. Faye O, Faye O, Diallo D, Diallo M, Weidmann M, Sall AA. Quantitative real-time PCR detection of Zika virus and evaluation with field-caught mosquitoes. Virol J. 2013;10:311.View ArticleGoogle Scholar
  23. Balm MND, Lee CK, Lee HK, Chiu L, Koay ESC, Tang JW. A diagnostic polymerase chain reaction assay for Zika virus. J Med Virol. 2012;84:1501–5.View ArticleGoogle Scholar
  24. Scaramozzino N, Crance J-M, Jouan A, DeBriel DA, Stoll F, Garin D. Comparison of flavivirus universal primer pairs and development of a rapid, highly sensitive heminested reverse transcription-PCR assay for detection of flaviviruses targeted to a conserved region of the NS5 gene sequences. J Clin Microbiol. 2001;39:1922–7.View ArticleGoogle Scholar
  25. Pierre V, Drouet MT, Deubel V. Identification of mosquito-borne flavivirus sequences using universal primers and reverse transcription/polymerase chain reaction. Res Virology. 1994;145:93–104.View ArticleGoogle Scholar
  26. Semenza JC, Zeller H. Integrated surveillance for prevention and control of emerging vector-borne diseases in Europe. Euro Surveill. 2013;54:2–5.Google Scholar
  27. Gourinat AC, O’Connor O, Calvez E, Goarant C, Dupont-Rouzeyrol M. Detection of Zika virus in urine. Emerg Infect Dis. 2015;21:84–6.View ArticleGoogle Scholar
  28. Musso D, Roche C, Nhan T-X, Robin E, Teissier A. Detection of Zika virus in saliva. J Clin Virol. 2015;68:53–5.View ArticleGoogle Scholar
  29. Atkinson B, Hearn P, Afrough B, Lumley S, Carter D, Aarons EJ, et al. Detection of Zika virus in semen. Emerg Infect Dis. 2016;22:940.View ArticleGoogle Scholar
  30. FDA (U.S. Food and Drug Administration). Emergency Use Authorizations: Zika Virus Emergency Use Authorization. FDA. 26 Feb 2016. http://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496.htm#zika. Accessed 20 Oct 2016.
  31. Luminex. xMAP® MultiFLEX™ ZIKA RNA ASSAY Product Details. Luminex Corp. 2016.Google Scholar
  32. Hologic. Aptima® Zika Virus Assay Package Insert. Hologic, Inc. 2016.Google Scholar
  33. Ahrberg CD, Manz A, Neužil P. Palm-sized device for point-of-care Ebola detection. Anal Chem. 2016;88:4803–7.View ArticleGoogle Scholar
  34. Lim S, Nan H, Lee MJ, Kang SH. Fast on-site diagnosis of influenza A virus by Palm PCR and portable capillary electrophoresis. J Chromatogr B. 2014;963:134–9.View ArticleGoogle Scholar
  35. Ahrberg CD, Ilic BR, Manz A, Neužil P. Handheld real-time PCR device. Lab Chip. 2016;16:586–92.View ArticleGoogle Scholar
  36. Li TJ, Chang CM, Chang PY, Chuang YC, Huang CC, Su WC, et al. Handheld energy-efficient magneto-optical real-time quantitative PCR device for target DNA enrichment and quantification. NPG Asia Mat. 2016;8:e277.View ArticleGoogle Scholar
  37. Harshman DK, Rao BM, McLain JE, Watts GS, Yoon JY. Innovative qPCR using interfacial effects to enable low threshold cycle detection and inhibition relief. Science Adv. 2015;1:e1400061.View ArticleGoogle Scholar
  38. Fakruddin M, Mannan KS, Chowdhury A, Mazumdar RM, Hossain MN, Islam S, Chowdhury MA. Nucleic acid amplification: alternative methods of polymerase chain reaction. J Pharm Bioall Sci. 2013;5:245–52.View ArticleGoogle Scholar
  39. Pardee K, Green AA, Takahashi MK, Braff D, Lambert G, Lee JW, et al. Rapid, low-cost detection of Zika virus using programmable biomolecular components. Cell. 2016;165:1255–66.View ArticleGoogle Scholar
  40. Song J, Mauk MG, Hackett BA, Cherry S, Bau HH, Liu C. Instrument-free point-of-care molecular detection of Zika virus. Anal Chem. 2016;88:7289–94.View ArticleGoogle Scholar
  41. Tian B, Qiu Z, Ma J, de la Torre TZG, Johansson C, Svedlindh P, Strömberg M. Attomolar Zika virus oligonucleotide detection based on loop-mediated isothermal amplification and AC susceptometry. Biosens Bioelectron. 2016;86:420–5.View ArticleGoogle Scholar
  42. Eboigbodin KE, Brummer M, Ojalehto T, Hoser M. Rapid molecular diagnostic test for Zika virus with low demands on sample preparation and instrumentation. Diag Microb Infect Dis. 2016. doi:https://doi.org/10.1016/j.diagmicrobio.2016.08.027.
  43. Curtis KA, Rudolph DL, Nejad I, Singleton J, Beddoe A, Weigl B, LaBarre P, Owen SM. Isothermal amplification using a chemical heating device for point-of-care detection of HIV-1. PLoS One. 2012;7:e31432.View ArticleGoogle Scholar
  44. Singleton J, Osborn JL, Lillis L, Hawkins K, Guelig D, Price W, Johns R, Ebels K, Boyle D, Weigl B, LaBarre P. Electricity-free amplification and detection for molecular point-of-care diagnosis of HIV-1. PLoS One. 2014;9:e113693.View ArticleGoogle Scholar
  45. Sayad AA, Ibrahim F, Uddin SM, Pei KX, Mohktar MS, Madou M, Thong KL. A microfluidic lab-on-a-disc integrated loop mediated isothermal amplification for foodborne pathogen detection. Sens Actuat B-Chem. 2016;227:600–9.View ArticleGoogle Scholar
  46. Jenkins DM, Kubota R, Dong J, Li Y, Higashiguchi D. Handheld device for real-time, quantitative, LAMP-based detection of Salmonella enterica using assimilating probes. Biosens Bioelectron. 2011;30:255–60.View ArticleGoogle Scholar
  47. Lu W, Wang J, Wu Q, Sun J, Chen Y, Zhang L, Zheng C, Gao W, Liu Y, Jiang X. High-throughput sample-to-answer detection of DNA/RNA in crude samples within functionalized micro-pipette tips. Biosens Bioelectron. 2016;75:28–33.View ArticleGoogle Scholar
  48. Craw P, Mackay RE, Naveenathayalan A, Hudson C, Branavan M, Sadiq ST, Balachandran W. A simple, low-cost platform for real-time isothermal nucleic acid amplification. Sensors. 2015;15:23418–30.View ArticleGoogle Scholar
  49. Liu C, Mauk MG, Hart R, Qiu X, Bau HH. A self-heating cartridge for molecular diagnostics. Lab Chip. 2011;11:2686–92.View ArticleGoogle Scholar
  50. Huang S, Do J, Mahalanabis M, Fan A, Zhao L, Jepeal L, Singh SK, Klapperich CM. Low cost extraction and isothermal amplification of DNA for infectious diarrhea diagnosis. PLoS One. 2013;8:e60059.View ArticleGoogle Scholar
  51. Iwamoto T, Sonobe T, Hayashi K. Loop-mediated isothermal amplification for direct detection of mycobacterium tuberculosis complex, M. avium, and M. intracellulare in sputum samples. J Clin Microbiol. 2003;41:2616–22.View ArticleGoogle Scholar
  52. Mori Y, Nagamine K, Tomita N, Notomi T. Detection of loop-mediated isothermal amplification reaction by turbidity derived from magnesium pyrophosphate formation. Biochem Biophys Res Comm. 2001;289:150–4.View ArticleGoogle Scholar
  53. Nagatani N, Yamanaka K, Saito M, Koketsu R, Sasaki T, Ikuta K, Miyahara T, Tamiya E. Semi-real time electrochemical monitoring for influenza virus RNA by reverse transcription loop-mediated isothermal amplification using a USB powered portable potentiostat. Analyst. 2011;136:5143–50.View ArticleGoogle Scholar
  54. Tomita N, Mori Y, Kanda H, Notomi T. Loop-mediated isothermal amplification (LAMP) of gene sequences and simple visual detection of products. Nat Protoc. 2008;3:877–82.View ArticleGoogle Scholar
  55. Ma XJ, Shu YL, Nie K, Qin M, Wang DY, Gao RB, Wang M, Wen LY, Han F, Zhou SM, Zhao X. Visual detection of pandemic influenza A H1N1 Virus 2009 by reverse-transcription loop-mediated isothermal amplification with hydroxynaphthol blue dye. J Virol Meth. 2010;167:214–7.View ArticleGoogle Scholar
  56. Stedtfeld RD, Stedtfeld TM, Kronlein M, Seyrig G, Steffan RJ, Cupples AM, Hashsham SA. DNA extraction-free quantification of Dehalococcoides spp. in groundwater using a hand-held device. Environ Sci Technol. 2014;48:13855–63.View ArticleGoogle Scholar
  57. Rodriguez-Manzano J, Karymov MA, Begolo S, Selck DA, Zhukov DV, Jue E, Ismagilov RF. Reading Out single-molecule digital RNA and DNA isothermal amplification in nanoliter volumes with unmodified camera phones. ACS Nano. 2016;10:3102–13.View ArticleGoogle Scholar
  58. Myers FB, Henrikson RH, Bone J, Lee LP. A handheld point-of-care genomic diagnostic system. PLoS One. 2013;8:e70266.View ArticleGoogle Scholar
  59. Zhao H, Fernandez E, Dowd KA, Speer SD, Platt DJ, Gorman MJ, et al. Structural basis of Zika virus-specific antibody protection. Cell. 2016;166:1016–27.View ArticleGoogle Scholar
  60. CDC. Zika MAC-ELISA. CDC. 29 June 2016. http://www.cdc.gov/zika/pdfs/zika-mac-elisa-instructions-for-use.pdf. Accessed 2 Oct 2016.
  61. Cecchetto J, Fernandes FCB, Lopes R, Bueno PR. The capacitive sensing of NS1 Flavivirus biomarker. Biosens Bioelectron. 2017;87:949–56.View ArticleGoogle Scholar
  62. Cavalcanti IT, Guedes MIF, Sotomayor MDPT, Yamanaka H, Dutra RF. A label-free immunosensor based on recordable compact disk chip for early diagnostic of the dengue virus infection. Biochem Eng J. 2012;67:225–30.View ArticleGoogle Scholar
  63. Jahanshahi P, Zalnezhad E, Sekaran SD, Adikan FRM. Rapid immunoglobulin M-based dengue diagnostic test using surface Plasmon resonance biosensor. Sci Rep. 2014;4:3851.View ArticleGoogle Scholar
  64. Gyurech D, Schilling J, Schmidt-Chanasit J, Cassinotti P, Kaeppeli F, Dobec M. False positive dengue NS1 antigen test in a traveler with an acute Zika virus infection imported into Switzerland. Swiss Med Wkly. 2016;146:w14296.Google Scholar
  65. Dai L, Song J, Lu X, Deng Y-Q, Musyoki AM, Cheng H, et al. Structures of the Zika virus envelope protein and its complex with a flavivirus broadly protective antibody. Cell Host Microbe. 2016;19:696–704.View ArticleGoogle Scholar
  66. Badawi MM, Osman MM, Alla AAF, Ahmedani AM, hamed Abdalla M, Gasemelseed MM, et al. Highly conserved epitopes of Zika envelope glycoprotein may act as a novel peptide vaccine with high coverage: immunoinformatics approach. Am J Biomed Res. 2016;4:46–60.Google Scholar
  67. Meltzer E, Leshem E, Lustig Y, Gottesman G, Schwartz E. The clinical spectrum of Zika virus in returning travelers. Am J Med. 2016;129:1126–30.View ArticleGoogle Scholar
  68. Acharya D, Bastola P, Le L, Paul AM, Fernandez E, Diamond MS, et al. An ultrasensitive electrogenerated chemiluminescence-based immunoassay for specific detection of Zika virus. Sci Rep. 2016;6:32227.View ArticleGoogle Scholar